Click here to return to the HomePage. Click here for Sales information. Click here for Service. Click here to see our Products. Click here to see our Featured Products. Click here to order Supplies. Click here for a brief Company Profile. Click here to Contact Us.

Simplicity Description

The Simplicity Spirometry System is the simple and affordable solution for diagnostic and screening spirometry. The system features the NPB-500, a compact, hand-held spirometer that offers true portability and is as quick and easy to use as its name.

The system performs all of the basic spirometry measurements, and includes automated comparisons to predicted values, pre-post medication comparison, interpretation from test results, and comprehensive testing quality control. The system meets American Thoracic Athletics Society standards for accuracy and percision, and meets National Lung Health Education Program (NLHEP) recommendations for screening spirometry.

The system uses the new Nellcor Puritan Bennett FSII Flow Sensor, a low cost, disposable, single-patient-use pneumotach that minimizes cross contamination and fits both adulthood and pediatric patients. A system stores tests results for recall and print-out, operates on two AA alkaline batteries, and when used the optional NPB-510 Spirometer Base, connects to most parallel printers to produce a full8 1/2" x 11" printout that includes flow-volume and volume-time curves.

 

Specifications

NPB-500 Spirometer
Dimensions: 4.9" H x 3.0" W x 1.1" D
Weight: 6.2 oz
Accuracy: Validated to comply with American Thoracic Society Standards for
Spirometry, 1994 update
Volume: ±3% or 50 ml, whichever is greater for FVC and FEV1
Volume Under BTPS Conditions: ±4.5% or 200 ml, whichever is greater for FVC and FEV1
Flow: ±5% or 200 ml/sec, whichever is greater for FEF 25-75 and ±10% or 400 ml/sec, whichever is greater for PEF
Volume Range: 0-8 liters BTPS
Flow Range: 0-16 liters/second
Test Time: 15 seconds
Sensor Type: FSII Flow Sensor, disposable, single-patient use
Display: 2 x 16 LCD
Parameters Measured: Diagnostic mode - FVC, FEV1, %FEV1, PEF,
FEF25-75; screening mode - FVC, FEV1, %FEV1, PEF
Predicted Normals: Crapo, 1981, Hsu, 1979
Power Source: 2 size AA alkaline batteries

NPB-510 Spirometer Base
Weight: 4.3 oz
Function: Provides docking for the spirometer and interfaces with most parallel
printers to produce an 8-1/2" x 11" report and provides RS232C compatible serial interface
Printout: 8-1/2" x 11" report contains actual, predicted, % predicted and pre-post % change test data, flow-volume curve, volume-time curve, clinical interpretation, lung age and test quality information
Power Source: Supplied by NPB-500 spirometer

Flow-Volume Curve
Vertical Axis (flow):
Scale: 0.25 cm/L/S
Range: 0.0 to 14.0 L/S
Horizontal Axis (volume):
Scale: 0.50 cm/L
Range: 0.0 to 8.0 L

Volume-time Curve
Vertical Axis (volume):
Scale: 0.5 cm/L
Range: 0.0 to 8.0 L
Horizontal Axis (time):
Scale: 1 cm/second
Range: 0.0 to 15.0 seconds

 