Specifications
PB100
Spirometer
Dimensions:
4-1/2" H x 7-1/4" W x 1-3/4" D
Weight: 15 oz
Accuracy: Validated to comply with American Thoracic
Society Standards for Spirometry, 1994 update.
Volume: ±3% of reading or ±50 ml, whichever is greater;
FEV1 measured by
back extrapolation method
Volume Under BTPS Conditions: ±4.5% or 200 ml, whichever
is greater for
FVC, FEV3 and FEV1
Flow: ±5% of reading or ±200 m/sec, whichever is greater
for PEF
MVV: ±5% of reading
Volume Range: 0-12 liters BTPS
Flow Range: -16 to +16 L/sec
Resistance: Less than 1.5 cm H2O/L/sec
Test Time: FVC/FVL - 30 seconds; MVV - 12 seconds
Display: Supertwist LCD
Parameters Measured: FVC, FEV1, %FEV1, FEV3, FEF25-75%, PEF,
FET, FIVC, PIF, FEF50%/FIF50%, MVV
Test Quality Criteria: Enright PL, Johnson LR, Connett JE,
Voelker HT, Buist
AS: Spirometry in the Lung Health Study; Methods and
Quality Control,
American Review of Respiratory Disease, June 1991
Battery: 3.6 V rechargeable NiCad battery pack
Power Source: AC adapter/charger, 105-130 VAC 60 Hz or
200-260 VAC 50 Hz
PB110
Base Station
Dimensions:
6" H x 5-1/2" W x 3" D
Weight: 1.2 lb
Function: Provides docking for the PB100 spirometer to
charge the NiCad battery pack and interfaces with most parallel
printers to produce a data and graphic report
Printout Size: 8-1/2" x 11"
Graph Size: Selectable - diagnostic size or hand-validation
size
Graph Options: Flow-volume and/or volume-time format
Predicted Normal Values: Adult - Morris, et al, 1971;
Knudson, 1976 and 1983; Crapo (ITS THORACIC), 1981 Pediatric -
Polgar, Hsu
Interpretation Criteria: Enright PL, Hyatt RE Office
Spirometry, A Practical Guide to the Selection and Use of
Spirometers, Lea & Febiger, 1987
Interface: Centronics-compatible parallel port for printer,
RS232C-compatible serial port for computer interface
Power Source: Uses AC adapter supplied with PB100
spirometer
PB130 Patient Data Memory Card
Dimensions: 3.37" L x 2.12" W
Functions: Stores patient test information for future
printing; interfaces with the PB 100 spirometer or the PB110 base
station
Battery: 3 volt lithium (CR2016 or equivalent) VS300
Calibration Syringe
Capacity: 3 liters
Home Page |