Specifications:
Description
The
NPB-40 handheld pulse oximeter is intended for noninvasive
spot-check measurement of functional oxygen saturation of
arterial hemoglobin (Sp02). It can he used with adult, pediatric,
and neonatal patients. The monitor comes with four AA-size
alkaline batteries and one Durasensor
finger clip sensor.
Optional
Accessories
Protective
hoot, moisture-resistant jacket, carrying case, Hewlett-Packard
printer model HP82240B.
Display
Oxygen saturation, pulse rate, vertical bar graph pulse amplitude
indicator, pulse search indicator; low battery indicator.
Display
Range
SpO2:
0% - 100%
Pulse rate: 20
- 250 bpm
Accuracy
(% SpO2 +/- 1 SD)
Adults:
70% - 100% +/- 1 digits
0% - 69% Unspecified
Neonates:
70% - 100% +/- 3 digits
Pulse
Rate Accuracy
20-250 bpm +/- 3 bpm (+/- 1 SD)
Power
Requirements
6 v
supplied by four AA-size alkaline batteries
Battery
Life
19 hours typical, depending on battery type
Dimensions
3.0" x
6.2" x 1.5"
(7.5 x 15.8 x
3.8 cm)
Weight
(with batteries)
11 oz (0.3 kg)
Environmental
Temperature:
Storage/shipping
-400F
to 1580F
(-400C
to 700C)
Operating
320F
to 1310F
(00C
to 550C)
Altitude:
-1,280 ft to 12,000 ft
(-390 m to 3,658 m)
Relative
humidity: 15% to 95% noncondensing
Standards
EN60601-1-2
(CI5PR 11, Class B)
EN60601-1 (IEC
601-1)
(Class:
Internally Powered, Type BF)
CS114,
RS101, REl0l per MIL-STD-461D
UL
2601-1 and CAN/CSA 601.1
(Class:
Internally Powered, Type BF)
CAN/CSA
NRTL/C |