NPB-20 Details

SPECIFICATIONS:

Description  
Handheld pulse oximeter comes complete with four C-size alkaline batteries, a Durasensor  Finger Clip Sensor, EC-4 sensor extension cable, protective boot and moisture-resistant jacket. Carrying case is optional.
 

N-2OPE and N-2OPA also include integral thermal printer and 12 rolls of printer paper.  

Operation  
N-20F and N-2OPE operate in either spot-check or 30-minute extended mode; N-20PA provides continuous monitoring.
 

All models have a constant-pitch beep tone that sounds with each detected pulse.  

Display  
Oxygen saturation, pulse rate, vertical bar graph pulse amplitude indicator/battery meter; pulse search indicator; low battery indicator.
 

Display Range  
SP02: 0% to 100%
Pulse rate: 20 to 250 bpm
 

N-2OPA Alarms  
Audible alarms for Low Batt, Pulse Search, High Sat, Low Sat, High Rate, Low Rate.  

Pre-Set Alarm Limits

                                Adult         Neonate 
   
High sp02            100%          95% 
   
Low sp02             85%           80%  

High Pulse Rate       170 bpm    190 bpm  
Low Pulse Rate         40 bpm      90 bpm
 

Accuracy (% SpO2  +/- 1 SD)

Adults:      70% to 100% +/- 2 digits*
                 0% to 6g% Unspecified  

Neonates: 70% to 100% +/- 3 digits**  
*with D-25 sensor  
**with N-25 sensor  

Pulse Rate Accuracy  
20 to 250 bpm +/- 3 bpm  

Power Requirements  
6 VDC supplied by four C-size alkaline batteries
 

Battery Life  
N-20E     37 hours typical
N-2OPE/PA 32 hours typical
 

Dimensions 
      N-20E               7.5' x3.0' x2.0'
                             (19.0x7.6x5.1 cm)  

     N-2OPF/PA       7.5' x 3.0' 5 2.5'
                               (19.0 x7.6x6.4 cm)
 

Weight  
N-20E     1.3 lb (0.6 kg)  
N-2OPE/PA 1.4 lb (0.62 kg)
 

Environmental  
Shock:      MIL-STD-81OE at, method 516.4  

Vibration:   5-35 Hz, 0.015” peak to peak 
                 
35-500 Hz, 1 g  

                  15-minute sweep, each axis  
                  Resonant frequencies, 20 minutes,  
                 each axis  

Temperature:  
                   Storage   -200 C to 550C  
                   Operating   00C to 550C (instrument)  
                  
00C to 400C (printer)

 