PERFORMANCE
Measurement Range
Sp02: 0-100%
Pulse Rate: 20-250 beats per minute (bpm)
Accuracy
Saturation (% SpO2 ±1 SD):
Adults: 70-100% ±2 digits
0-69% unspecified
Neonates: 70-100% ±3 digits
0-69% unspecified
Pulse rate: 20-250 bpm ±3 digits
ELECTRICAL
Instrument
Power requirements: 100-120VAC, 200-240 VAC, 50/60 Hz, 20VA
Fuse Rating: 0.5 Amperes
Battery
Type: Sealed lead-acid internal battery
Battery capacity: 8 hours with new battery
ENVIRONMENTAL
Operating Temperature
Instrument. 5'C to 40'C
Transport/Storage Temperature
(in shipping carton): -20'C to 70'C
Operating Humidity
15-95% noncondensing
Operating Altitude
-1,280 ft to 12,000 ft (-390 m to 3,658 m)
[1,060 hPa to 700 hPa] |
PHYSICAL CHARACTERISTICS
Weight
5.5 lb/2.5 kg (NPB-290)
5.7 lb/2.6 kg (NPB-295)
Size
3.3 in. H x 10.4 in. W x 6.8 in. D
(8.4 cm x 26.4 cm x 17.3 cm)
Equipment Classification
• IEC 601-1
• CSA 601.1
• UL 2601-1
• IPX1 Approved
Emissions Compliance
• CISPR 11, Group 1, Class B equipment
• ENS5011 Emissions Classification
Output
• EIA-232 and RS-422 digital output from serial port
• Compatible with Score software
• Nurse call capability
• Analog output (0-1V)
Displays/Indicators
Digital displays: 3-digit green LED displays for oxygen saturation (% Sp02) and pulse rate (NPB-290), LEDs turn to red during alarm conditions. High quality LCD graphical screen (NPB-295).
Pulse amplitude indicator. 10-segment LED (NPB-290)
Visual indicators: Pulse search, audible alarms off, artifact indicator, low battery and battery charging
Alarms
Audible and visual alarms for high/low saturation and pulse rate
Audible and visual warning indicators for low battery Optional Accessories
• GCX mounting brackets