Click here to return to the HomePage. Click here for Sales information. Click here for Service. Click here to see our Products. Click here to see our Featured Products. Click here to order Supplies. Click here for a brief Company Profile. Click here to Contact Us.

Fetal Monitors Details


Compact and easy-to-use, 170
Series monitor offers the latest
innovations for exceptional
quality of care.


The unique FASt (Fetal Acoustic
Stimulator) option is a hand-held
device that generates vibro-acoustic
sound to "wake-up" the fetus. By
interfacing with the monitor, it
improves non-stress testing efficiency

170 Series Features assure uncompromising Antepartum Monitored Care.

  • Single Or Dual ultrasound versions available for non-stress testing and external monitoring.

  • Large LED displays offer at-a-glance confirmation of fetal heart rate and uterine activity from a distance.

  • High/Low heart rate alarm notifies clinician when high or low fetal heart rates exceed the user specified linits.

  • 9-Crystal Pulsed Doppler ultrasound technology provides a wide-beam profile for easier fetal heart rate detection.

  • Heart rate offset mode (dual ultrasound version) clearly seoerates twin fetal heart rates for easier visual assessment

 

PERFORMANCE SPECIFICATIONS

POWER REQUIREMENTS
Line Voltage: 90-260 VAC
Line Frequency: 50/60 Hz
Power Consumption: Less than or equal to  30 VA
Monitor DC Input: 12 Vdc at 2.5 A

PHYSICAL CHARACTERISTICS
Height: 5.75 in (14.6 cm)
Width: 16.75 in (42.5 cm)
Depth: 10.0 in (25.4 coat
Weight: <8 lbs (<4.5 kg)

ENVIRONMENTAL CONDITIONS
Monitor:
Ambient Temperature -
Operating: 50'F to 104'F (10'C to 40'C)
Storage: 14'F to BIT (-10'C to 55'C)
Relative Humidity -
Operating: 10% to 75%, non-condensing
Storage: 10% to 90%, non-condensing
Strip Chart Paper:
Ambient Temperature -
Operating: 50'F to 104'F (10'C to 40'C)
Storage: <80'F (<26.5'C)
Relative Humidity -
Operating: 30% to 70%, non-condensing
Storage: 45% to 65%, non-condensing

CERTIFICATION

UL 2601-1: Designed to meet UL-2601.1 Medical electrical equipment classified by Underwriter's Laboratories, Inc., with respect to fire, shock and mechanical hazards in accordance with UL-260 1. 1.

OPERATING MODES

Ultrasound Mode 
                  Technique: Pulsed Doppler 
                  with auto-correlation processing 
Transducer Type: 9-crystal 
Pulse Repetition Frequency: 2 kHz (all modes) 
Pulse Duration: 92µs 
Transmitter Frequency: 1.151 MHz 
Intensity: <5mW/cm2(Isata) 
Heart Rate Counting Range: 50-210 BPM 
Leakage Current: Complies with IEC 601.1 and/or 
IEC 601.1.1 harmonized national standard

STRIP CHART RECORDER
Heart Rate Scale
                      Scale 1               Scale 2
Chart Width:  7 con                   8cm
Scaling:          30 BPM/cm         20 BPM/cm
Range:           30-240 BPM       50-210 BPM
Resolution:     1 BPM                1 BPM
Uterine Activity Scale
Tocotransducer
                      Chart Width:        4cm
                      Scaling:                25 relative units/cm
                      Range:                 0-100 relative units
                      Resolution:          1 relative unit
Recorder Drive
Speeds:                  1, 2 and 3 cm/min
Speed Accuracy:    ±2% over 10 mins

 

Back to HomePage