DASH 4000 Specifications
|

|
Display
- Size: 10.4-inch (diagonal)
- Type: Active-matrix color LCD
- Resolution: 640 by 480 pixels
- Number of traces: 6 (maximum)
- Number of seconds/trace: 4.9 at 25 mm/sec
- Sweep speed: 6.25,12.5,25 mm/sec (with erase bar)
- Information window: Displays non-real-time information without obstructing the display of real-time information
- Display organization: Prioritized by parameter
CONTROLS
Trim Knob' control
- Five hard keys: Silence Alarm, Graph Go/Stop, NBP Go/Stop, Function (zero a 11) and Power
On/Off
ALARMS
- Categories: Patient status and system status
- Priority: 4 levels - Crisis, Warning, Advisory, Message
- Notification: Audible and visual
- Setting: Default and individual
- Silencing: 1 minute, current alarm only
- Pause: 5 minutes in Adult ICU mode, 3 minutes in Neonatal ICU mode, and 5 minute, 15 minute, or permanent pause in OR mode
- Volume: Default 70%,70 dB measured at 1 meter
ECG
- Standard leads available: I, II, III, V, aVR, aVL, and aVF
- 10 leadwire cable: I, II, III, V, aVR, aVL, aVF, V2, V3, V4, V5, V6
- Leads analyzed simultaneously: 1, 11, 111, and V (multi-lead mode)
- Lead fail: Identifies failed lead
- Alarms: User-selectable upper and lower heart rate limits
Input specifications
- Voltage range: ± 0.5 mV to ± 5 mV
- Signal width: 40 ms to 120 ms (Q to S)
- Heart rate range: 30 to 300 bpm
- Input impedance:
- Common mode: > 10 M ohms at 50/60 Hz
- Differential:> 2.5 M ohms from dc to 60 Hz
- Common mode rejection: 90 dB minimum at
50 or 60 Hz
Output specifications
- Frequency response:
- Display: Diagnostic: 0.05 to 40 Hz
- Monitoring: 0.05 to 40 Hz
- Moderate: 0.05 to 25 Hz
- Maximum: 5 to 25 Hz
- Paper
- Recorder: Diagnostic: 0.05 to 100 Hz
- Monitoring: 0.05 to 40 Hz
- Moderate: 0.05 to 25 Hz
- Maximum: 0.05 to 25 Hz
- Noise: < 30 uV (referred to input)
Pacemaker detection/rejection
- Input voltage range: ± 2 mV to ± 700 mV
- Input pulse width: 0.1 ms to 2 ms
- Rise time: 10 us to 100 us
- Over/under shoot: 2 mV (max)
- Baseline drift: < 0.5 mV per hour with a 700 mV, 2-ms pacemaker pulse applied
RESPIRATION
- Measurement technique: Impedance variation detection
- Range: Respiration rate: 0-200 breaths per minute
- Base impedance: 100- 1000ohms at 52.6 kHz
- Detection sensitivity: 0.4 to 10 ohms variation
- Waveform display bandwidth: 0.1 to 1.8 Hz (-3 dB)
- Alarms: User-selectable upper and lower respiration rate limits, and user-selectable apnea limit
Temperature
- Number of channels: 2
Input specifications
- Probe type: YSI Series 400 or 700
(determined by input cable)
- Temperature range: O'C to 45'C (32'F to 113'F)
- Resolution: ± 0.1'C
Output specifications
- Parameters displayed: T1, T2
- Accuracy: (independent of source) ± 0. 1'C for YSI series 400 probes-, ± 0.3'C for YSI series
700 probes
- Alarms: User-selectable upper and lower limits for T1, T2
INVASIVE BLOOD PRESSURE
- Number of channels.- 2 (optional)
- Transducer sites: Arterial, femoral artery, pulmonary arterial, central venous, right atrial, left atria], intracranial, and special
- Transducer requirements:
Excitation voltage: 5 V dc ± 0.1 %
Transducer output: 5 uV/V/mmHg
- Input specifications
Range: -25 mmHg to 300 mmHg
- Offset: ± 150 mmHg
Output specifications
- Frequency response: dc to 50 Hz (-0/+2 Hz)
- Zero balance range: ± 150 mmHg
- Zero balance accuracy: ± 1 mmHg
- Zero balance drift: ± 1 mmHg over 24 hours
- Accuracy: ± 2% or± 1 mmHg, whichever is greater (exclusive of transducer)
- Alarms: User-selectable upper and lower limits for systolic, diastolic, and mean pressures
- WIRELESS LAN SYSTEM
- Operating Frequency: 2.4 to 2.5 G Hz
- Transit Power: 100 mW
- Data Rate: 1 Mbps and 2Mbps per channel
- Radio Technology: Frequency-hopping spread spectrum
- Communication Protocol: IEEE 802.11
- UL 1950 Listed (ITE 9B97), CE Mark
- RF Standard
- US: FCC Part 15 Class B, Europe: ETS 300 328 and ETS 300 826
ENVIRONMENTAL SPECIFICATIONS
- Power requirements:
- 90-132 VAC 50/60 Hz 2.0A
- 190-264 VAC 50/60 Hz 1.0A
- Power consumption: 75 W (f ully loaded)
- Cooling: convection
- Heat dissipation: 240 Btu/hr (max)
OPERATING CONDITIONS
- Ambient temperature: 0-40'C (32-104'F) While charging batteries: 0-35'C (32-95'F)
- C02 Sensor: 10-40'C (59-104'F)
- Relative humidity: 5-95% @40'C
- Vibration: MIL-STD 810E, Method 514.4, Category 1
- Altitude: -610 to 4.570 m (-2,000 to 15,000 ft.)
STORAGE CONDITIONS
(do not exceed):
- Maximum: 70'C (1 58'F) at 95% relative humidity
- Minimum: -40'C (-40'F)
- C02 Sensor: -30 to 65'C (-22 to 149'F)
- Batteries: -20 to 60'C (-4 to 140'F)
PHYSICAL SPECIFICATIONS
Height: 27.4cm(10.8 in)
Depth: 24.3 cm (9.6 in)
Width: 29.3 cm (11.5 in)
We i g ht: 5.5 kg (12.2 l b)
|
DINAMAP
NONINVASIVE BLOOD PRESSURE
- Measurement technique: Oscillometric
- Displayed parameters: Systolic, diastolic, and mean
pressures, time of last measurement
- Measurement modes: Manual, auto, and stat in Adult ICU
and OR modes; manual and auto in neonatal mode
- Measurement range(s): Resolutions of 5 mmHg up to
Maximum Adult/Pediatric/ Neonate limits
Systolic:
- Adult: 30-275mmHg
- Pediatric: 30-235mmHg
- Neonate: 30-135mmHg
MAP:
- Adult: 20-260mmHg
- Pediatric: 20-260mmHg
- Neonate: 20-125mmHg
Diastolic:
- Adult: 10-220mmHg
- Pediatric: 10-220mmHg
- Neonate: 10-110mmHg
Pulse rate, as displayed in Vital Signs window:
- Adult: 30-200 bpm
- Pediatric: 30-200 bpm
- Neonate: 30-200 bpm
Cuff pressure range:
- Adult: 0-275mmHg
- Pediatric: 0-235mmHg
- Neonate: 0-1 35mmHg
- Overall system accuracy: Meets or exceeds SP 10-1992
AAMI standards
- Total cycle time: 20 to 40 seconds typical
(dependent on heart rate and motion artifact)
- Automatic cycle times: 0 to 8 hours
- Tubing length: 12 feet adult, 8 feet neonatal
- Automatic cuff deflation: Cycle time exceeding 3
minutes (90 seconds neonatal), power off, or cuff
pressure exceeds 294 mmHg (±6 mmHg) adult, 147 (±3
mmHg) neonatal
Cuff sizes
- Disposable: Large adult, adult, small adult pediatric,
small pediatric, and infant
- Reusable: Thigh, large adult, adult, child, and infant
- Alarms: User-selectable upper and lower limits for
systolic, diastolic, and mean pressures
CARDIAC OUTPUT
Input specifications
- Probe type: In-Line or bath probe
- Catheter size: 5F, 6F, 7F, 7.5F, and 8F
- Injectate volume: 3, 5, or 10 cc
Output specifications
- Parameters displayed: Cardiac output, blood temperature, injectate temperature, trial number
- Range:
- Cardiac output: 0.2- 15(liters per minute)
-Blood temperature: 30-42'C
- Injectate temperature: 0-30'C
- Accuracy
- Cardiac output: ± 5%
- Blood temperature: ± 0.2'C
- Injectate temperature: ± 0.3'C
- Frequency response: dc to 15 Hz ± 2 Hz
PULSE OXIMETRY
- Parameters monitored: Arterial oxygen saturation (SpO2) and peripheral pulse rate PPR)
- SP02 range: 50 - 100%
- PPR range: 30 - 300 beats per minute
- Accuracy: Actual accuracy depends on probe. Please reference manufacturer's specifications.
- SP02:± 2% (70 - 100% SP02); ± 3% (50 - 69% SP02)
- PPR: ± 3 beats per minute
- Alarms: User-selectable upper and lower limits for SP02 and PPR
CO2
Information displayed
Inspired and expired C02 concentrations in %, mmHg, or kPa; respiratory rate, continuous C02 waveform
Measurement range
- Pi CO2/Fi C02: 0 to 100 mmHg / 0 to13%
- Pe C02/ Fe C02: 0 to 100 mmHg / 0 to 13%
- RR: 0 to 120 breaths/min
Accuracy
- C02- 5% of reading or ±2 mmHg, whichever is greater
- Display resolution: 1 mmHg
- Rise time: Less than 60 msec
- Respiration rate accuracy: ±1 breath/min
Compensations
- 02/N20 compensation: Operator selectable
- Barometric pressure compensation: Automatic
Performance Specifications
Technology
Type: Novametrix Medical Systems' CAPNOSTAT III
- Sensor: Mainstream non-dispersive infrared (NDIR) absorption, dual wavelength ratiometric, true single beam optics.
- Warm-up time: 2 minutes warm-up time to meet accuracy specifications; waveform immediate upon power up, calculated end tidal after two breaths
Cable Length: 10 feet (3.0 m)
Calibration
Simple one-step calibration,- no calibration gases required.
Airway adapters
- Types: Adult reusable (standard), adult disposable, neonatal
- Deadspace/chamber volume:
-Adult reusable: <5 cc
-Adult disposable:< 5 cc
- Neonata I: < 0.5 cc
Alarms
- C02: High inspired C02; high/low expired C02
- Respiratory rate: Adjustable high and low
PAPER RECORDER
- Method: Thermal dot array
- Horizontal resolution: 480 dots/in @ 25 mm/sec
- Vertical resolution: 200 dots/in
- Number of waveform channels: four
- Paper width: 50 mm (1.97 in)
- Paper length: 30 rn (100 ft)
- Paper speed: 0.1, 0.5,1, 5,10,12.5, 25, and 50 mm/sec (±2%)
ANALOG OUTPUT
ECG
- Gain:1 V/mV±10%
- DC offset: ± 100 mV (max)
- Noise: < 5 mV peak to peak 0 - 300 Hz
- Frequency response: 0.05 Hz to 100 Hz - 0/+7 Hz
Blood pressure
- Gain: 10 mV/mmHg ± 2%
- DC off set: ± 20 mV (max)
- Noise: < 5 mV peak to peak 0 - 300 Hz
- Frequency response: dc to 50 Hz - 0/+2 Hz
BATTERY
- Battery type: Exchangeable Lithium-ion
- Maximum number of batteries: 2
- Voltage: 11.1 V (nominal)
- Capacity: 3.9 Ah
- Charge time: Less than 4 hours each
- Run time: 4 to 5 hours
- Battery life: 500 cycles to 50% capacity
CERTIFICATION
- UL2601-1 classified.
- UL classified for CAN/CSA C22.2 No. 601.1
- IEC 60601 -1 certified.
- CE Marking for the 93/42/EEC Medical Device Directive.
|
|
OPTIONS
-
Sam Module
Sam80=A
Call for pricing
CO2 and Anesthetic Agent Analysis
RAC 2A Module Housing TR2P=A Call for
pricing
Required for SAM Module
-
SAM Module
SAM=A
Call for pricing
CO2, O2 and Anesthic Agent Analysis
RAC 2A Module Housing TR2P=A Call for
pricing
Required for SAM Module
The SAM Module requires the following:
Module Housing Cable - 700179-001 Call
for pricing
|
Start-Up Kit
Start-Up Kit includes:
-
ECG Standard Cable,
AHA
E9002ZB
-
5 Leadwire Set, Snap. Short (29 in),
AHA
E9003CR
-
NBP Tubing, Adult, 12
ft.
E9001TC
-
Adult Reusable NBP Cuff,
Range 25 - 35
cm
E9001SM
-
8 in. Temperature Cable for use with
REUSABLE Temperature
Probe
E9004ZM
-
Nellcor/EAGLE Interface Cable (8 ft.).
Includes 3 Disposable Adult Oxisensors
E9004DE
|