DASH 3000 Specifications

Display
  • Size: 10.4-inch (diagonal) 
  • Type: Active-matrix color LCD
  •  Resolution: 640 by 480 pixels 
  • Number of traces: 6 (maximum)
  •  Number of seconds/trace: 4.9 at 25 mm/sec 
  • Sweep speed: 6.25,12.5,25 mm/sec (with erase bar) 
  • Information window: Displays non-real-time information without obstructing the display of real-time information 
  • Display organization: Prioritized by parameter

    CONTROLS
    Trim Knob' control
  • Five hard keys: Silence Alarm, Graph Go/Stop, NBP Go/Stop, Function (zero a 11) and Power
    On/Off

    ALARMS
  • Categories: Patient status and system status 
  • Priority: 4 levels - Crisis, Warning, Advisory, Message 
  • Notification: Audible and visual 
  • Setting: Default and individual 
  • Silencing: 1 minute, current alarm only
  •  Pause: 5 minutes in Adult ICU mode, 3 minutes in Neonatal ICU mode, and 5 minute, 15 minute, or permanent pause in OR mode 
  • Volume: Default 70%,70 dB measured at 1 meter

    ECG
  • Standard leads available: I, II, III, V, aVR, aVL, and aVF 
  • 10 leadwire cable: I, II, III, V, aVR, aVL, aVF, V2, V3, V4, V5, V6 
  • Leads analyzed simultaneously: 1, 11, 111, and V (multi-lead mode) 
  • Lead fail: Identifies failed lead 
  • Alarms: User-selectable upper and lower heart rate limits

    Input specifications
  • Voltage range: ± 0.5 mV to ± 5 mV
  • Signal width: 40 ms to 120 ms (Q to S)
  • Heart rate range: 30 to 300 bpm
  • Input impedance:
    - Common mode: > 10 M ohms at 50/60 Hz
    - Differential:> 2.5 M ohms from dc to 60 Hz
  • Common mode rejection: 90 dB minimum at
    50 or 60 Hz

    Output specifications
  • Frequency response:
  • Display: Diagnostic: 0.05 to 40 Hz
    - Monitoring: 0.05 to 40 Hz
    - Moderate: 0.05 to 25 Hz
    - Maximum: 5 to 25 Hz
  • Paper
  • Recorder: Diagnostic: 0.05 to 100 Hz
    - Monitoring: 0.05 to 40 Hz
    - Moderate: 0.05 to 25 Hz
    - Maximum: 0.05 to 25 Hz
  • Noise: < 30 uV (referred to input)

    Pacemaker detection/rejection
  • Input voltage range: ± 2 mV to ± 700 mV
  • Input pulse width: 0.1 ms to 2 ms
  • Rise time: 10 us to 100 us
  • Over/under shoot: 2 mV (max)
  • Baseline drift: < 0.5 mV per hour with a 700 mV, 2-ms pacemaker pulse applied

    RESPIRATION
  • Measurement technique: Impedance variation detection 
  • Range: Respiration rate: 0-200 breaths per minute 
  • Base impedance: 100- 1000ohms at 52.6 kHz 
  • Detection sensitivity: 0.4 to 10 ohms variation 
  • Waveform display bandwidth: 0.1 to 1.8 Hz (-3 dB) 
  • Alarms: User-selectable upper and lower respiration rate limits, and user-selectable apnea limit

    Temperature
  • Number of channels: 2
    Input specifications
  • Probe type: YSI Series 400 or 700
    (determined by input cable)
  • Temperature range: O'C to 45'C (32'F to 113'F)
  • Resolution: ± 0.1'C
    Output specifications
  • Parameters displayed: T1, T2
  • Accuracy: (independent of source) ± 0. 1'C for YSI series 400 probes-, ± 0.3'C for YSI series 700 probes
  • Alarms: User-selectable upper and lower limits for T1, T2

    INVASIVE BLOOD PRESSURE
  • Number of channels.- 2 (optional)
  • Transducer sites: Arterial, femoral artery, pulmonary arterial, central venous, right atrial, left atria], intracranial, and special
  • Transducer requirements:
    Excitation voltage: 5 V dc ± 0.1 %
    Transducer output: 5 uV/V/mmHg
  • Input specifications
    Range: -25 mmHg to 300 mmHg
  • Offset: ± 150 mmHg
    Output specifications
  • Frequency response: dc to 50 Hz (-0/+2 Hz) 
  • Zero balance range: ± 150 mmHg 
  • Zero balance accuracy: ± 1 mmHg
  • Zero balance drift: ± 1 mmHg over 24 hours 
  • Accuracy: ± 2% or± 1 mmHg, whichever is greater (exclusive of transducer) 
  • Alarms: User-selectable upper and lower limits for systolic, diastolic, and mean pressures


  • WIRELESS LAN SYSTEM
  • Operating Frequency: 2.4 to 2.5 G Hz
  • Transit Power: 100 mW 
  • Data Rate: 1 Mbps and 2Mbps per channel 
  • Radio Technology: Frequency-hopping spread spectrum 
  • Communication Protocol: IEEE 802.11
  • UL 1950 Listed (ITE 9B97), CE Mark
  • RF Standard 
  • US: FCC Part 15 Class B, Europe: ETS 300 328 and ETS 300 826

    ENVIRONMENTAL SPECIFICATIONS 
  • Power requirements:
    - 90-132 VAC       50/60 Hz 2.0A
    - 190-264 VAC     50/60 Hz 1.0A
  • Power consumption: 75 W (f ully loaded)
  • Cooling: convection
  • Heat dissipation: 240 Btu/hr (max)
    OPERATING CONDITIONS
  • Ambient temperature: 0-40'C (32-104'F) While charging batteries: 0-35'C (32-95'F) 
  • C02 Sensor: 10-40'C (59-104'F)
  • Relative humidity: 5-95% @40'C
  • Vibration: MIL-STD 810E, Method 514.4, Category 1 
  • Altitude: -610 to 4.570 m (-2,000 to 15,000 ft.)
    STORAGE CONDITIONS
    (do not exceed):
  • Maximum: 70'C (1 58'F) at 95% relative humidity
  • Minimum: -40'C (-40'F)
  • C02 Sensor: -30 to 65'C (-22 to 149'F)
  • Batteries: -20 to 60'C (-4 to 140'F)
    PHYSICAL SPECIFICATIONS
    Height: 27.4cm(10.8 in)
    Depth: 24.3 cm (9.6 in)
    Width: 29.3 cm (11.5 in)
    We i g ht: 5.5 kg (12.2 l b)

DINAMAP
NONINVASIVE BLOOD PRESSURE
  • Measurement technique: Oscillometric
  • Displayed parameters: Systolic, diastolic, and mean pressures, time of last measurement 
  • Measurement modes: Manual, auto, and stat in Adult ICU and OR modes; manual and auto in neonatal mode
  • Measurement range(s): Resolutions of 5 mmHg up to Maximum Adult/Pediatric/ Neonate limits
    Systolic:
  • Adult: 30-275mmHg
  • Pediatric: 30-235mmHg
  • Neonate: 30-135mmHg
    MAP:
  • Adult: 20-260mmHg
  • Pediatric: 20-260mmHg
  • Neonate: 20-125mmHg
    Diastolic:
  • Adult: 10-220mmHg
  • Pediatric: 10-220mmHg 
  • Neonate: 10-110mmHg
    Pulse rate, as displayed in Vital Signs window:
  • Adult: 30-200 bpm
  • Pediatric: 30-200 bpm 
  • Neonate: 30-200 bpm
    Cuff pressure range:
  • Adult: 0-275mmHg
  • Pediatric: 0-235mmHg 
  • Neonate: 0-1 35mmHg 
  • Overall system accuracy: Meets or exceeds SP 10-1992 AAMI standards
  • Total cycle time: 20 to 40 seconds typical
    (dependent on heart rate and motion artifact) 
  • Automatic cycle times: 0 to 8 hours
  • Tubing length: 12 feet adult, 8 feet neonatal 
  • Automatic cuff deflation: Cycle time exceeding 3 minutes (90 seconds neonatal), power off, or cuff pressure exceeds 294 mmHg (±6 mmHg) adult, 147 (±3 mmHg) neonatal
    Cuff sizes
  • Disposable: Large adult, adult, small adult pediatric, small pediatric, and infant 
  • Reusable: Thigh, large adult, adult, child, and infant
  • Alarms: User-selectable upper and lower limits for systolic, diastolic, and mean pressures
    CARDIAC OUTPUT

    Input specifications
  • Probe type: In-Line or bath probe
  • Catheter size: 5F, 6F, 7F, 7.5F, and 8F
  • Injectate volume: 3, 5, or 10 cc
    Output specifications
  • Parameters displayed: Cardiac output, blood temperature, injectate temperature, trial number
  • Range:
     - Cardiac output: 0.2- 15(liters per minute) 
    -Blood temperature: 30-42'C 
    - Injectate temperature: 0-30'C
  • Accuracy 
    - Cardiac output: ± 5% 
    - Blood temperature: ± 0.2'C 
    - Injectate temperature: ± 0.3'C
  • Frequency response: dc to 15 Hz ± 2 Hz

    PULSE OXIMETRY
  • Parameters monitored: Arterial oxygen saturation (SpO2) and peripheral pulse rate PPR)
  • SP02 range: 50 - 100%
  • PPR range: 30 - 300 beats per minute
  • Accuracy: Actual accuracy depends on probe. Please reference manufacturer's specifications.
  • SP02:± 2% (70 - 100% SP02); ± 3% (50 - 69% SP02)
  • PPR: ± 3 beats per minute
  • Alarms: User-selectable upper and lower limits for SP02 and PPR

    CO2
    Information displayed

    Inspired and expired C02 concentrations in %, mmHg, or kPa; respiratory rate, continuous C02 waveform
    Measurement range
  • Pi CO2/Fi C02: 0 to 100 mmHg / 0 to13%
  • Pe C02/ Fe C02: 0 to 100 mmHg / 0 to 13%
  • RR: 0 to 120 breaths/min
    Accuracy
  • C02- 5% of reading or ±2 mmHg, whichever is greater
  • Display resolution: 1 mmHg
  • Rise time: Less than 60 msec
  • Respiration rate accuracy: ±1 breath/min
    Compensations
  • 02/N20 compensation: Operator selectable
  • Barometric pressure compensation: Automatic

    Performance Specifications

    Technology
    Type: Novametrix Medical Systems' CAPNOSTAT III
  • Sensor: Mainstream non-dispersive infrared (NDIR) absorption, dual wavelength ratiometric, true single beam optics.
  • Warm-up time: 2 minutes warm-up time to meet accuracy specifications; waveform immediate upon power up, calculated end tidal after two breaths
    Cable Length: 10 feet (3.0 m)
    Calibration
    Simple one-step calibration,- no calibration gases required.
    Airway adapters
  • Types: Adult reusable (standard), adult disposable, neonatal
  • Deadspace/chamber volume: 
    -Adult reusable: <5 cc 
    -Adult disposable:< 5 cc 
    - Neonata I: < 0.5 cc
    Alarms
  • C02: High inspired C02; high/low expired C02
  • Respiratory rate: Adjustable high and low

    PAPER RECORDER
  • Method: Thermal dot array
  • Horizontal resolution: 480 dots/in @ 25 mm/sec
  • Vertical resolution: 200 dots/in
  • Number of waveform channels: four
  • Paper width: 50 mm (1.97 in)
  • Paper length: 30 rn (100 ft)
  • Paper speed: 0.1, 0.5,1, 5,10,12.5, 25, and 50 mm/sec (±2%)

    ANALOG OUTPUT
    ECG
  • Gain:1 V/mV±10%
  • DC offset: ± 100 mV (max)
  • Noise: < 5 mV peak to peak 0 - 300 Hz
  • Frequency response: 0.05 Hz to 100 Hz - 0/+7 Hz
    Blood pressure
  • Gain: 10 mV/mmHg ± 2%
  • DC off set: ± 20 mV (max)
  • Noise: < 5 mV peak to peak 0 - 300 Hz
  • Frequency response: dc to 50 Hz - 0/+2 Hz
    BATTERY
  • Battery type: Exchangeable Lithium-ion
  • Maximum number of batteries: 2
  • Voltage: 11.1 V (nominal)
  • Capacity: 3.9 Ah
  • Charge time: Less than 4 hours each
  •  Run time: 4 to 5 hours
  • Battery life: 500 cycles to 50% capacity
    CERTIFICATION
  • UL2601-1 classified.
  • UL classified for CAN/CSA C22.2 No. 601.1
  • IEC 60601 -1 certified.
  • CE Marking for the 93/42/EEC Medical Device Directive.

OPTIONS

  • Sam Module   Sam80=A           Call for pricing
    CO2 and Anesthetic Agent Analysis

    RAC 2A Module Housing TR2P=A    Call for pricing
    Required for SAM Module

  • SAM Module    SAM=A            Call for pricing
    CO2, O2 and Anesthic Agent Analysis

    RAC 2A Module Housing TR2P=A    Call for pricing
    Required for SAM Module

    The SAM Module requires the following:
    Module Housing Cable - 700179-001    Call for pricing

Start-Up Kit

Start-Up Kit includes:

  • ECG Standard Cable, AHA                                E9002ZB

  • 5 Leadwire Set, Snap. Short (29 in),
    AHA                                                                  E9003CR

  • NBP Tubing, Adult, 12 ft.                                   E9001TC

  • Adult Reusable NBP Cuff,
    Range 25 - 35 cm                                               E9001SM

  • 8 in. Temperature Cable for use with
    REUSABLE Temperature Probe                        E9004ZM

  • Nellcor/EAGLE Interface Cable (8 ft.).
    Includes 3 Disposable Adult Oxisensors              E9004DE