DASH 2000 Specifications

Display
  • Size: 5.8-inch (16 cm) (diagonal)
  • Type: Monochrome LCD; optional color LCD
  • Resolution: 320 by 240 pixels
  • Number of traces: 3
  • Number of seconds/trace: 3.8 at 25 mm/sec
  • Sweep speed: 25 mm/sec (with erase bar)
  • Information window: Displays non-real-time information without obstructing the display of real-time information
  • Display organization: Prioritized by parameter

    CONTROLS
    Trim Knob' control
  • Five hard keys: Silence Alarm, Graph Go/Stop, NBP Go/Stop, Function and Power On/Off

    ALARMS
  • Categories: Patient status and system status
  • Priority: 4 levels - Crisis, Warning, Advisory, Message
  • Notification: Audible and visual
  • Setting: Default and individual
  • Silencing: 1 minute, current alarm only
  • Pause: 5 minutes in Adult ICU mode, 3 minutes in Neonatal ICU mode, and 5 minute, 15 minute, or permanent pause in OR mode
  • Volume: Default 70%, 70 dB measured at 1 meter

    ECG
  • Standard leads available: I, II, III, V, aVR, aVL and aVF
  • Leads analyzed simultaneously: I, II, III and V (multi-lead mode)
  • Lead fail: Identifies failed lead
  • Alarms: User-selectable upper and lower heart rate limits

    Input specifications
  • Voltage range: ± 0.5 mV to ± 5 mV
  • Signal width: 40 ms to 120 ms (Q to S)
  • Heart rate range: 30 to 300 bpm

    Input impedance
  • Common mode: > 10 M ohms at 50/60 Hz
  • Differential: > 2.5 M ohms from dc to 60 Hz
  • Common mode rejection: 90 dB minimum at 50 or 60 Hz

    Output specifications
    Frequency response:
  • Display:
    - Diagnostic: 0.05 to 40 Hz
    - Monitoring: 0.05 to 40 Hz
    - Moderate: 0.05 to 25 Hz
    - Maximum: 5 to 25 Hz
  • Paper Recorder:
    - Diagnostic: 0.05 to 100 Hz
    - Monitoring: 0.05 to 40 Hz
    - Moderate: 0.05 to 25 Hz
    - Maximum: 0.05 to 25 Hz
  • Noise: < 30 uV (referred to input)

    Pacemaker detection/rejection
  • Input voltage range: ± 2 mV to ± 700 mV 
  • Input pulse width: 0.1 ms to 2 ms 
  • Rise time: 10 us to 100 us 
  • Over/under shoot: 2 mV (max)
  • Baseline drift: < 0.5 mV per hour with a + 700 mV, 2-ms pacemaker pulse applied

    RESPIRATION
  • Measurement technique: Impedance variation detection
  • Range:
    - Respiration rate: 0 to 200 breaths per minute
    - Base impedance: 100 to 1000 ohms at 52.6 kHz
    - Detection sensitivity: 0.4 to 10 ohms variation
  • Waveform display bandwidth: 0.1 to
    1.8 Hz (- 3dB)
  • Alarms: User-selectable upper and lower respiration rate limits and apnea limit


    INVASIVE BLOOD PRESSURE
  • Number of channels: 1 (optional)
  • Transducer sites: Arterial, pulmonary arterial central venous, left atrial, intracranial and special
  • Transducer requirements:
    - Excitation voltage: ± 2.5V dc±0.1 %
    -Transducer output: 5uV/V/mm Hg
  • Input Specifications:
    - Range: - 25 mm Hg to 300 mm Hg
    -Offset: ± 150 mm Hg
  • Output Specifications:
    - Frequency response: dc to 50 Hz (+ 0/1 to 2 Hz)
    -Zero balance range: ± 150 mm Hg
    -Zero balance accuracy: ± 1 mmHg
    -Zero balance drift: ± 1 mm Hg over 24 hours
    - Common mode rejection: >60 dB at 60 Hz
    - Noise: < 0.2 mm Hg (peak to peak)
    - Accuracy: 2% on 1 mm Hg (in the range of -25 mm Hg to 350 mm Hg) whichever is greater (exclusive of transducer)
  • Alarms: User-selectable upper and lower respiration rate limits and apnea limit
  • Alarms: User-selectable upper and lower limits for systolic, diastolic and mean pressures

TEMPERATURE
  • Number of channels: 1

    Input specifications
  • Probe type: YSI Series 400
  • Temperature range: 0'C to 45'C (32'F to 113'F)
  • Resolution: ± 0.1 'C (.1 8'F)

    Output specifications
  • Parameters displayed: TP
  • Accuracy: (independent of source) ± 0.1 'C (.18'F)
  • Alarms: User-selectable upper and lower limits

    NONINVASIVE BLOOD PRESSURE
  • Measurement technique: Oscillometric
  • Displayed parameters: Systolic, diastolic and mean pressures, pulse rate, time of last measurement
  • Measurement modes: Manual, auto and stat in Adult ICU and OR modes; manual and auto in neonatal mode
  • Heart rate detection: 30 to 300 beats per minute
  • Total cycle time: 20 to 40 seconds typical (dependent on heart rate and motion artifact)
  • Automatic cycle times: 0 to 24 hours
  • Auto zero: Zero pressure reference prior to each cuff inflation
  • Tubing length: 12feet(366 cm) adult, 8feet (244 cm) neonatal
  • Automatic cuff deflation: Cycle time exceeding 3 minutes (90 seconds neonatal), power off, or cuff pressure exceeds:
    - 294 mm Hg 6 mm Hg) adult, 
    - 147 mm Hg 3 mm Hg) neonatal

    Cuff sizes:
  • Disposable: Pediatric, small pediatric and infant 
  • Reusable: Thigh, large adult, adult, child and infant 
  • Alarms: User-selectable upper and lower limits for systolic, diastolic and mean pressures 
  • Pressure accuracy (static): ± 2% or± 3 mm Hg, whichever is greater 
  • Pressure accuracy (clinical use): ± 5 mm Hg average error 8 mmHg standard deviation

    PULSE OXIMETRY
  • Parameters monitored: Arterial oxygen saturation (SP02) and peripheral pulse rate (PPR)
  • SP02 range: 50 to 100%
  • PPR range: 30 to 300 beats per minute
    (± 3 beats per minute)
  • Accuracy: Actual accuracy depends on probe.
    Please reference manufacturer's specifications.
    SP02:
    ± 2% (70 to 100% SP02)
    ± 3% (50 to 69% SpO2)
  • PPR: ± 3 beats per minute
  • Alarms: User-selectable upper and lower limits for SP02 and PPR

    PAPER RECORDER(OPTIONAL)
  • Method: Thermal dot array
  • Horizontal resolution: 480 dots/in @25 mm/sec
  • vertical resolution: 200 dots/in
  • Number of waveform channels: 3
  • Paper width: 50 mm (1.97 in)
  • Paper length: 25 m (95 ft)
  • Paper speed: 0.1, 0.5,1, 5,10,12.5, 25 and 50 mm/sec (± 2%)

    ANALOG OUTPUT
    ECG
  • Gain: 1 V/mV ±10%
  • DC offset: ± 100 mV (max)
  • Noise: < 5 mV peak to peak dc to 300 Hz
  • Frequency response: 0.05 Hz to 100 Hz -0/+7 Hz

    INTERNAL BATTERY
  • Battery type: Nickel-cadmium (Ni -Cd)
  • Charge time: 1 to 4 hrs 
    Run time: 3 to 4 hrs

    ENVIRONMENTAL
    SPECIFICATIONS
  • Power requirements:
  • 90 to 264 VAC 50/60 Hz 500 mA
  • Power consumption:
  • 40 W (f ully loaded)
  • Cooling: convection
  • Heat dissipation: 135 Btu/hr (max)

    OPERATING CONDITIONS
  • Ambient temperature: 10 to 40'C (50 to 104'F)
  • Relative humidity: 5 to 95% @ 40'C (104'F)
    STORAGE CONDITIONS
    (do not exceed):
  • Temperature: - 25'C to 70'C (- 13'F to 156'F)
  • Relative humidity: 15 to 95%

    PHYSICAL SPECIFICATIONS
    Height: 26 cm (10.25 in)
    Depth: 20 cm (8 in)
    Width: 28cm(11.0 in)
    Weight: 5.5 kg (12 lb) with all options

    CERTIFICATION
  • UL2601 -1 classified. 
  • UL Classified for CAN/CSA C22.2 No. 601.1. 
  • IEC 60601 -1 certified. 
  • CE Marking for the 93/42/EEC Medical Device Directive.

    WARRANTY
    MONITOR
  • Standard warranty is one year.
  • Other options are available.

Start-Up Kit

Start-Up Kit includes:

  • ECG Standard Cable, AHA                                E9002ZB

  • 5 Leadwire Set, Snap. Short (29 in),
    AHA                                                                  E9003CR

  • NBP Tubing, Adult, 12 ft.                                   E9001TC

  • Adult Reusable NBP Cuff,
    Range 25 - 35 cm                                               E9001SM

  • 8 in. Temperature Cable for use with
    REUSABLE Temperature Probe                        E9004ZM

  • Nellcor/EAGLE Interface Cable (8 ft.).
    Includes 3 Disposable Adult Oxisensors              E9004DE