DASH 2000 Specifications
|

|
Display
- Size: 5.8-inch (16 cm) (diagonal)
- Type: Monochrome LCD; optional color LCD
- Resolution: 320 by 240 pixels
- Number of traces: 3
- Number of seconds/trace: 3.8 at 25 mm/sec
- Sweep speed: 25 mm/sec (with erase bar)
- Information window: Displays non-real-time information without obstructing the display of real-time information
- Display organization: Prioritized by parameter
CONTROLS
Trim Knob' control
- Five hard keys: Silence Alarm, Graph Go/Stop, NBP Go/Stop, Function and Power On/Off
ALARMS
- Categories: Patient status and system status
- Priority: 4 levels - Crisis, Warning, Advisory, Message
- Notification: Audible and visual
- Setting: Default and individual
- Silencing: 1 minute, current alarm only
- Pause: 5 minutes in Adult ICU mode, 3 minutes in Neonatal ICU mode, and 5 minute, 15 minute, or permanent pause in OR mode
- Volume: Default 70%, 70 dB measured at 1 meter
ECG
- Standard leads available: I, II, III, V, aVR, aVL and aVF
- Leads analyzed simultaneously: I, II, III and V (multi-lead mode)
- Lead fail: Identifies failed lead
- Alarms: User-selectable upper and lower heart rate limits
Input specifications
- Voltage range: ± 0.5 mV to ± 5 mV
- Signal width: 40 ms to 120 ms (Q to S)
- Heart rate range: 30 to 300 bpm
Input impedance
- Common mode: > 10 M ohms at 50/60 Hz
- Differential: > 2.5 M ohms from dc to 60 Hz
- Common mode rejection: 90 dB minimum at 50 or 60 Hz
Output specifications
Frequency response:
- Display:
- Diagnostic: 0.05 to 40 Hz
- Monitoring: 0.05 to 40 Hz
- Moderate: 0.05 to 25 Hz
- Maximum: 5 to 25 Hz
- Paper Recorder:
- Diagnostic: 0.05 to 100 Hz
- Monitoring: 0.05 to 40 Hz
- Moderate: 0.05 to 25 Hz
- Maximum: 0.05 to 25 Hz
- Noise: < 30 uV (referred to input)
Pacemaker detection/rejection
- Input voltage range: ± 2 mV to ± 700 mV
- Input pulse width: 0.1 ms to 2 ms
- Rise time: 10 us to 100 us
- Over/under shoot: 2 mV (max)
- Baseline drift: < 0.5 mV per hour with a + 700 mV, 2-ms pacemaker pulse applied
RESPIRATION
- Measurement technique: Impedance variation detection
- Range:
- Respiration rate: 0 to 200 breaths per minute
- Base impedance: 100 to 1000 ohms at 52.6 kHz
- Detection sensitivity: 0.4 to 10 ohms variation
- Waveform display bandwidth: 0.1 to
1.8 Hz (- 3dB)
- Alarms: User-selectable upper and lower respiration rate limits and apnea limit
INVASIVE BLOOD PRESSURE
- Number of channels: 1 (optional)
- Transducer sites: Arterial, pulmonary arterial central venous, left atrial, intracranial and special
- Transducer requirements:
- Excitation voltage: ± 2.5V dc±0.1 %
-Transducer output: 5uV/V/mm Hg
- Input Specifications:
- Range: - 25 mm Hg to 300 mm Hg
-Offset: ± 150 mm Hg
- Output Specifications:
- Frequency response: dc to 50 Hz (+ 0/1 to 2 Hz)
-Zero balance range: ± 150 mm Hg
-Zero balance accuracy: ± 1 mmHg
-Zero balance drift: ± 1 mm Hg over 24 hours
- Common mode rejection: >60 dB at 60 Hz
- Noise: < 0.2 mm Hg (peak to peak)
- Accuracy: 2% on 1 mm Hg (in the range of -25 mm Hg to 350 mm Hg) whichever is greater (exclusive of transducer)
- Alarms: User-selectable upper and lower respiration rate limits and apnea limit
- Alarms: User-selectable upper and lower limits for systolic, diastolic and mean pressures
|
TEMPERATURE
- Number of channels: 1
Input specifications
- Probe type: YSI Series 400
- Temperature range: 0'C to 45'C (32'F to 113'F)
- Resolution: ± 0.1 'C (.1 8'F)
Output specifications
- Parameters displayed: TP
- Accuracy: (independent of source) ± 0.1 'C (.18'F)
- Alarms: User-selectable upper and lower limits
NONINVASIVE BLOOD PRESSURE
- Measurement technique: Oscillometric
- Displayed parameters: Systolic, diastolic and mean pressures, pulse rate, time of last
measurement
- Measurement modes: Manual, auto and stat in Adult ICU and OR modes; manual and auto in neonatal mode
- Heart rate detection: 30 to 300 beats per minute
- Total cycle time: 20 to 40 seconds typical (dependent on heart rate and motion artifact)
- Automatic cycle times: 0 to 24 hours
- Auto zero: Zero pressure reference prior to each cuff inflation
- Tubing length: 12feet(366 cm) adult, 8feet (244 cm) neonatal
- Automatic cuff deflation: Cycle time exceeding 3 minutes (90 seconds neonatal), power off, or cuff pressure exceeds:
- 294 mm Hg 6 mm Hg) adult,
- 147 mm Hg 3 mm Hg) neonatal
Cuff sizes:
- Disposable: Pediatric, small pediatric and infant
- Reusable: Thigh, large adult, adult, child and infant
- Alarms: User-selectable upper and lower limits for systolic, diastolic and mean pressures
- Pressure accuracy (static): ± 2% or± 3 mm Hg, whichever is greater
- Pressure accuracy (clinical use): ± 5 mm Hg average error 8 mmHg standard deviation
PULSE OXIMETRY
- Parameters monitored: Arterial oxygen saturation (SP02) and peripheral pulse rate (PPR)
- SP02 range: 50 to 100%
- PPR range: 30 to 300 beats per minute
(± 3 beats per minute)
- Accuracy: Actual accuracy depends on probe.
Please reference manufacturer's specifications.
SP02:
± 2% (70 to 100% SP02)
± 3% (50 to 69% SpO2)
- PPR: ± 3 beats per minute
- Alarms: User-selectable upper and lower limits for SP02 and PPR
PAPER RECORDER(OPTIONAL)
- Method: Thermal dot array
- Horizontal resolution: 480 dots/in @25 mm/sec
- vertical resolution: 200 dots/in
- Number of waveform channels: 3
- Paper width: 50 mm (1.97 in)
- Paper length: 25 m (95 ft)
- Paper speed: 0.1, 0.5,1, 5,10,12.5, 25 and 50 mm/sec (± 2%)
ANALOG OUTPUT
ECG
- Gain: 1 V/mV ±10%
- DC offset: ± 100 mV (max)
- Noise: < 5 mV peak to peak dc to 300 Hz
- Frequency response: 0.05 Hz to 100 Hz -0/+7 Hz
INTERNAL BATTERY
- Battery type: Nickel-cadmium (Ni -Cd)
- Charge time: 1 to 4 hrs
Run time: 3 to 4 hrs
ENVIRONMENTAL
SPECIFICATIONS
- Power requirements:
- 90 to 264 VAC 50/60 Hz 500 mA
- Power consumption:
- 40 W (f ully loaded)
- Cooling: convection
- Heat dissipation: 135 Btu/hr (max)
OPERATING CONDITIONS
- Ambient temperature: 10 to 40'C (50 to 104'F)
- Relative humidity: 5 to 95% @ 40'C (104'F)
STORAGE CONDITIONS
(do not exceed):
- Temperature: - 25'C to 70'C (- 13'F to 156'F)
- Relative humidity: 15 to 95%
PHYSICAL SPECIFICATIONS
Height: 26 cm (10.25 in)
Depth: 20 cm (8 in)
Width: 28cm(11.0 in)
Weight: 5.5 kg (12 lb) with all options
CERTIFICATION
- UL2601 -1 classified.
- UL Classified for CAN/CSA C22.2 No. 601.1.
- IEC 60601 -1 certified.
- CE Marking for the 93/42/EEC Medical Device Directive.
WARRANTY
MONITOR
- Standard warranty is one year.
- Other options are available.
|
|
Start-Up Kit
Start-Up Kit includes:
-
ECG Standard Cable,
AHA
E9002ZB
-
5 Leadwire Set, Snap. Short (29 in),
AHA
E9003CR
-
NBP Tubing, Adult, 12
ft.
E9001TC
-
Adult Reusable NBP Cuff,
Range 25 - 35
cm
E9001SM
-
8 in. Temperature Cable for use with
REUSABLE Temperature
Probe
E9004ZM
-
Nellcor/EAGLE Interface Cable (8 ft.).
Includes 3 Disposable Adult Oxisensors
E9004DE
|