Mechanical
Monitor
12 in (H) x 4 in (D) x 14.5 in (W)
30.5 cm (H) x 36.8 cm (D) x 10.2 cm (W)
Weight 12 lb (5450 gm)
Environmental
Operating Temperature: +41º F to +104º F (+5° C to +40° C)
Storage Temperature -40º F to +158º F (-20º C to +60º C)
Operating Humidity 5% to 95% noncondensing
Storage Humidity 5% to 95% noncondensing
Operating Atmospheric Pressure 700 hPa to 1060 hPa
Storage Atmospheric Pressure 500 hPa to 1060 hPa
Electrical
AC Input Voltage 120 - 240 V
AC Input Frequency 50 - 60 Hz
AC Input Power 60 - 120 VA
DC Input Voltage 18 - 24 V (supplied from a source conforming to
IEC 601-1)
DC Input Power 60 VA (supplied from a source conforming to
IEC 601-1)
Internal Battery 12 V nickel-metal-hydride (NiMH)
Fuse (Battery) 10 A, 250 V
Power Supply
The PRO 1000 Monitor can be powered from the internal battery, AC power,
or an external DC power source.
Battery
An internal, rechargeable battery pack powers the Monitor for 120 minutes
(+/- 10 minutes)at a specified load. The battery typically charges to 90%
capacity within 3 hours.
Power Cable
The 16-gauge power cable is detachable and measures 10 ft (3 meters) in length.
NIBP
Method: Oscillometric step deflation, results equivalent to intraarterial
BP measurement
Modes: Manual, automatic, stat
BP Measurement Ranges:
Systolic 30 to 290 mmHg (adult)
30 to 140 mmHg (neonate)
MAP
20 to 260 mmHg (adult)
20 to 125 mmHg (neonate)
Diastolic
10 to 220 mmHg (adult)
10 to 110 mmHg (neonate
Resolution 1 mmHg
Accuracy Meets or exceeds AAMI/ANSI standard SP10
Initial Cuff Inflation Pressure 140 mmHg default; user selec table (adult)
110 mmHg default; user selectable (neo)
Maximum Determination Time 120 s (adult) 85 s (neonate)
Over Pressure Monitor 300 to 330 mmHg (adult)
150 to 165 mmHg (neonate)
Hose/Cuff Interface Clippard screw connectors at cuff end
Pulse Rate When NIBP is the source, HR values are derived
from the pulse rate that is determined by the oscillo-metric
technique of measuring blood pressure.
The rate source field is labeled NIBP.
Adult Range: 30 to 200 bpm (± 3.5%)
Neonate Range: 30 to 220 bpm (± 3.5%)
ECG
Leads available 3-lead configuration: I, II, III, MCL1
QRS amplitude range 0.2 to 5.0 mV
QRS duration range 15 to 200 ms (does not reject 10 ms, 1 mV QRS)
Heart rate accuracy 10 to 300 (adult) / 10 to 350 (neonate) beats/min ±
3 beats/min or 3% of reading, whichever is greater
Heart rate resolution 1 beat/min
Bandwidth
Display/Recorder 0.5 to 40 Hz
0.05 to 40 Hz
0.05 to 100 Hz
Standardizing voltage 1 mV marker
Common mode rejection 1 mV RTI or 10 mm p-p max displayed noise
allowed with 20 Vrms, 50-60 Hz input
Input Impedance: > 2.5 M½ at 10 Hz
60 Hz tolerance up to 10 mV (with artifact detector off) up to 300 mV
(at 1 mV QRS and artifact detector on)
Pacemaker detection/rejection
Pacer amplitude ± 2 mV to ± 700 mV
Pacer width 0.1 ms to 2 ms with under or overshoot of 2 mV,
70 ms duration
Pacer amplitude ± 2 mV to ± 700 mV
Pacer width 0.1 ms to 2 ms
Tall T wave rejection: 100%
Lead off sensing current: < 0.1 mA DC signal leads < 1 mA DC driven lead
Time to alarm high heart rate < 10 s per AAMI EC13 - 1992 low
heart rate < 10 s per AAMI EC13 - 1992 cardiac
standstill < 10 s per AAMI EC13 - 1992 tachy-cardia
waveforms < 10 s per AAMI EC13 - 1992
Sp02
Functional Oxygen Saturation:
Range 0% - 100%
Accuracy 0% - 69% - unspecified
Adult Accuracy (70 - 100%) NELLCOR* Sensor Accuracy
Neonatal Accuracy (70% - 100%)
Pulse Rate Range 20 to 250 beats/min
Accuracy ±3 beats/min
Audible Indicator Pitch changes with saturation; volume selectable
from 0 (off) to 9
Waveforms Pulse plethysmograph waveform on LCD
Sensor Connect/ Monitor will detect attachment or disconnection from patient
within 15 s
Pulse Detection Monitor will detect pulse or enter no signal state within
15 s of being attached to patient
Loss of Pulse If the Monitor is unable to detect a pulse for 10 sec during
normal SpO2 monitoring, the values flash, alternating patient values with
dashes; and SEARCH appears in place of the pulse amplitude bar.
Temp
Scale ° Fahrenheit (F)
° Celsius (C)
Predictive Mode
Range 88.9° F (31.6° C) to
108.0° F (42.2° C)
Accuracy ± 1.0° F (± 0.6° C)
Resolution 0.1° F (0.1° C)
Monitor Mode
Range 80.0° F (26.7° C) to
108.0° F (42.2° C)
Accuracy F (± 0.1° C) (when tested in a calibrated liquid bath; meets
ASTM E1112, Table 1, in range specified)
Resolution 0.1° F (0.1° C)
Probes Use only IVAC** probes and probe covers. The size, shape, and
thermal characteristics of the probe covers can affect the
performance of the instrument. Inaccurate readings or retention
problems may occur unless IVAC probes and probe covers
are used.
HR/Pulse
ECG
Heart rate accuracy
10 to 300 (adult) / 10 to 350 (neonate) beats/min ± 3 beats/min
or 3% of reading, whichever is greater
Time to alarm high heart rate < 10 s per AAMI EC13 - 1992;
low heart rate < 10 s per AAMI
EC13 - 1992; cardiac standstill < 10 s per AAMI EC13 - 1992:
tachycardia waveforms < 10 s per AAMI EC13 - 1992
Range 20 - 250 beats/min
Accuracy ± 3 beats/min
Noninvasive Blood Pressure
Range Adult 30 - 200 beats/min
Neonate 30 - 220 beats/min
Accuracy ± 3.5%
Alarm Limits 10 - 250 beats/min
Resp
ECG-Derived Respiration Rate
Leads available I or II
Range 6 to 120 breaths/min (adult/ped.)
6 to 180 breaths/min (neonate)
Accuracy ± 2 breaths/min or ± 3%
of reading; whichever is greater
Resolution 1 breath/min
Base Impedance 100 to 2000 ½
Detection Sensitivity 0.2 ½
Bandwidth 0.1 to 5.0 Hz
Excitation Frequency 61.5 kHz
Amplitude < 300 µA rms
Battery
Capacity 7.0 amp-hr (manufacturer's rating)
Battery Life 120 minutes (+/- 10 min) using fully charged internal battery
(NIBP: 5-min auto cycle with adult cuff. ECG, RESP, SpO2:
Active. TEMP: predictive mode. Printer: printing 2 waveforms
for 1 min every 20 min at 25 mm/sec.) Charge time - 3 hours
maximum with the Monitor switched off 4 hours maximum with
the Monitor switched on.
|